eLINE Electronic Pipettes
Brand : Biohit
The electronic Biohit eLINE pipette is the market leader in performance and ergonomics. This award winning (Pro Finnish Design 2001) pipette takes the effort out of pipetting and significantly reduces the risk of Repetitive Strain Injury (RSI).
The comprehensive range of pipetting programs in eLINE reduces the need for several work stages and enables you to dispense liquids twice as fast compared to a mechanical pipette. The unique DC-motor concept with build-in error control improves pipetting precision and allows more reliable results.
eLINE is available in single-channel models covering the volume range of 0.1 to 5000 μl and multichannel models from 0.2 to 1200 μl.
Description
he electronic Biohit eLINE pipette is the market leader in performance and ergonomics. This award winning (Pro Finnish Design 2001) pipette takes the effort out of pipetting and significantly reduces the risk of Repetitive Strain Injury (RSI).
The comprehensive range of pipetting programs in eLINE reduces the need for several work stages and enables you to dispense liquids twice as fast compared to a mechanical pipette. The unique DC-motor concept with build-in error control improves pipetting precision and allows more reliable results.
eLINE is available in single-channel models covering the volume range of 0.1 to 5000 μl and multichannel models from 0.2 to 1200 μl.
Key benefits:
- Excellent accuracy and precision with unique DC motor concept
- Fully electronic operations guarantee user-independent results and prevent RSI
- Easiest tip ejection ever with the electronic tip ejection
- Twice as fast as a mechanical pipette
- One eLINE covers the volume range of two mechanical pipettes – less pipettes needed
- Comprehensive range of pipetting programs
- Several charging options
Features
- Unbeatable accuracy and precision with unique DC motor concept
- Fully electronic operations guarantee user-independent results
- Patented and built-in error control for more reliable results
- The easiest tip ejection ever with the unique electronic tip ejection
- Ergonomic design that suits both left- and right-handed
- Optiload mechanism, originally developed by Biohit, in multichannel models for perfect tip sealing
- Easy programming, 6 memory places for favorite programs
- Different charging options (charging stand, charging carousel and direct charging)
- Single point calibration for special liquids
- Autoclavable dispensing heads (excl.1200 µl 8/12 channel models)
- Safe Cone Filters to prevent contamination and damage of pipette
- 2-year warranty
- CE/IVD marked
Operational Modes
- Pipetting
- Reverse Pipetting
- Super Pipetting *
- Pipetting+Mixing
- Multiple Dispensing
- Diluting
- Diluting+Mixing
- Sequential Dispening **
- Multi Aspirating
* Only in eLINE 0.1-5 µl
** Excl. eLINE 0.1-5 µl
Specification
Item | Channels | Volume | Increment | Test volume | Inaccuracy | Imprecision |
eLINE Electronic Pipette | 1 | 0.2-10 µl | 0.05 µl | 10 µl 5 µl 1 µl 0.2 µl |
0.90 % 1.00 % 2.50 % 12.00 % |
0.40 % 0.70 % 1.50 % 10.00 % |
eLINE Electronic Pipette | 1 | 5-120 µl | 0.50 µl | 120 µl 60 µl 12 µl 5 µl |
0.40 % 0.60 % 2.00 % 3.50 % |
0.15 % 0.20 % 1.00 % 1.50 % |
eLINE Electronic Pipette | 1 | 10-300 µl | 1.00 µl | 300 µl 150 µl 30 µl 10 µl |
0.40 % 0.60 % 1.50 % 3.50 % |
0.15 % 0.20 % 0.80 % 2.00 % |
eLINE Electronic Pipette | 1 | 50-1000 µl | 5.00 µl | 1000 µl 500 µl 100 µl 50 µl |
0.40 % 0.60 % 1.50 % 2.00 % |
0.15 % 0.20 % 0.50 % 1.00 % |
eLINE Electronic Pipette | 1 | 100-5000 µl | 10.0 µl | 5000 µl 2500 µl 500 µl |
0.50 % 0.80 % 1.00 % |
0.15 % 0.20 % 0.40 % |
eLINE Electronic Pipette | 1 | 0.1-5 µl | 0.05 µl | 5.0 µl 2.5 µl 0.5 µl |
1.20 % 1.70 % 9.50 % |
0.70 % 1.20 % 8.00 % |
eLINE Electronic Pipette | 8 | 0.2-10 µl | 0.05 µl | 10 µl 5 µl 1 µl |
0.90 % 1.50 % 4.00 % |
0.50 % 0.80 % 3.00 % |
eLINE Electronic Pipette | 8 | 5-120 µl | 0.50 µl | 120 µl 60 µl 12 µl |
0.50 % 0.70 % 2.00 % |
0.20 % 0.30 % 1.50 % |
eLINE Electronic Pipette | 8 | 10-300 µl | 1.00 µl | 300 µl 150 µl 30 µl |
0.50 % 0.70 % 2.00 % |
0.20 % 0.30 % 1.00 % |
eLINE Electronic Pipette | 8 | 50-1200 µl | 5.00 µl | 1200 µl 600 µl 120 µl |
0.50 % 1.00 % 2.50 % |
0.20 % 0.30 % 1.00 % |
eLINE Electronic Pipette | 12 | 0.2-10 µl | 0.05 µl | 10 µl 5 µl 1 µl |
0.90 % 1.50 % 4.00 % |
0.50 % 0.80 % 3.00 % |
eLINE Electronic Pipette | 12 | 5-120 µl | 0.50 µl | 120 µl 60 µl 12 µl |
0.50 % 0.70 % 2.00 % |
0.20 % 0.30 % 1.50 % |
eLINE Electronic Pipette | 12 | 10-300 µl | 1.00 µl | 300 µl 150 µl 30 µl |
0.50 % 0.70 % 2.00 % |
0.20 % 0.30 % 1.00 % |
eLINE Electronic Pipette | 12 | 50-1200 µl | 5.00 µl | 1200 µl 600 µl 120 µl |
0.50 % 1.00 % 2.50 % |
0.20 % 0.30 % 1.00 % |
* Note: Minimum volume in Pipetting mode is 500µl. The 100 µl is possible in Multiple Dispensing mode.