MA150
MA150
Brand : Sartorius
Brand : Sartorius
Weighing Capacity: 150g. Weighing accuracy: 1 mg
Readability : 0.01%. Temperature range: 40˚C-220˚C
Some Features
Multiple heating modes. Multiple heating units selection Hinged, flip-up sample cover. Optional version compliant with FDA|HACCP Regulations
Description
The moisture meter MA150 is synonymous with easy operation and a time-tested, space saving housing design for routine use in the laboratory, incoming inspection and in-process control.
Some highlighted features of MA150
- 150g weighing capacity with 1mg resolutions
- Multiple weighing elements available; ceramic or CQR (coiled quartz radiator)
- Fully automatic endpoint shutoff mode when the sample the sample reaches a constant weigh.
- Memory for 20 moisture measurement programs
- FDA/HACCP-compliant models (without glass parts) available.
Specification
Heating element |
Ceramic heating, gentle and smooth
|
Capacity |
150 g
|
Number of measuring programs |
20
|
Application |
Thermo-sensitive organic samples
|
Recommended moisture range |
1 - 100%
|
Readability |
1 mg, 0,01 %
|
Line voltage |
230 V
|
Max. weighing capacity |
150 g
|
Accuracy of the weighing system |
1 mg
|
Weighing system with EC type-approval certificate |
No
|
for initial sample weight approx. 1 g |
+/- 0,2
|
for initial sample weight, approx. 5 g |
+/- 0,05
|
Readability |
0.01 %
|
Display mode for results |
% moisture, % dry weight (solids), % RATIO, g residue
|
Heating mode |
Standard drying, gentle drying
|
Analysis mode |
Fully automatic, Semi-automatic
|
asap mode for parameter recognition |
No
|
Heating unit (choice of) |
Ceramic IR heating element, CQR heater (coiled quartz radiator)
|
Exchange of the heating unit by Plug&Dry® |
No
|
Access to the sample chamber |
Via hinged cover
|
Version compliant with FDA/ HACCP regulations |
Yes, all versions with the appendix "V2"
|
Built-in calibration weight |
No
|
Operator guidance features |
Context-sensitive menu with alphanumeric interactive prompts and symbols, parameter input using soft-key prompts
|
reproTEST |
No
|
Number of program memories |
20
|
Memory data storage |
End point up to the next moisture measurement run
|
Parameter settings password-protection |
Yes
|
Manual input of tare weights |
No
|
Data Printer |
External (optional)
|
Data interface |
RS232C unidirectional
|
Bar code scanner can be connected |
No
|
In-use dust cover for keypad |
Yes
|
Width |
213 mm
|
Depth |
320 mm
|
Height |
180,5 mm
|
Weight (Approx.) |
5,5 kg
|
ISO/GLP compliant print out |
Inalterable standard configuration template
|
Temperature range and settings |
40°C - 180°C, adjustable in 1-degree increments
|